Agilent RapidFire High-Throughput Mass Spectrometry System Application Note

This application note describes a rapid method for detecting benzoylecgonine (BZE), the major metabolite of cocaine, in urine using the Agilent RapidFire High-throughput Mass Spectrometry System. Results were achieved with much faster sample cycle times and similar analytical output compared to GC/MS or LC/MS assays.


Agilent RapidFire High-Throughput Mass Spectrometry System Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
463 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
548A06C44F659A1E5D45C1DA750263C4

About this manual

This document presents a study evaluating the Agilent RapidFire High-Throughput Mass Spectrometry System for analyzing benzoylecgonine (BZE), the major metabolite of cocaine, in urine samples. The system is capable of analyzing samples with cycle times of less than 15 seconds, offering high throughput compared to traditional GC/MS or LC/MS assays.

The study details the experimental setup, including the use of an Agilent RapidFire 360 coupled with an Agilent 6460 Triple Quadrupole Mass Spectrometer. It outlines the specific chemical conditions, sample preparation methods involving 96-well plates, and data analysis procedures using MassHunter and RapidFire Integrator software. Results demonstrate accurate and precise measurement of BZE with a limit of detection less than 2 ng/mL and a throughput greater than 240 samples per hour.

What's inside

  • Introduction
  • Experimental
  • RapidFire triple quadrupole conditions
  • Chemicals and reagents
  • Sample preparation
  • Data analysis
  • Results and Discussion
  • Conclusions

Specifications

Sample cycle timeless than 15 seconds
Throughputgreater than 240 samples per hour
Limit of detection (LOD)less than 2 ng/mL
Injection volume10 pL
Gas temperature350 °C
Sheath gas temperature400 °C
Nozzle voltage300 V
Capillary voltage3,500 V

Frequently asked questions

What is the sample cycle time for the Agilent RapidFire system?

The system is capable of analyzing samples with cycle times of less than 15 seconds.

What is the limit of detection for BZE in this method?

The limit of detection (LOD) for BZE was determined to be less than 2 ng/mL.

What is the maximum throughput of the system?

The methodology is capable of throughputs greater than 240 samples per hour.