Agilent 7700x ICP-MS Application Note for Food Sample Analysis

To ensure food safety and safeguard human health, the characterization of the elemental composition of a wide range of food types is required. The possibility of measuring all the required elements with a single configuration using the Agilent 7700x ICP-MS is investigated. The determination of trace and major elements simultaneously is possible due to the 9 orders of magnitude provided by the detector. Collision cell technology with a single cell gas, helium, is used to eliminate interferences generated by the plasma and the sample matrix. The use of a discrete sampling system to increase sample throughput is also evaluated.


Agilent 7700x ICP-MS Application Note for Food Sample Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
453 KB
Published
11 June, 2012
Updated
28 September, 2023
Model
7700x
MD5 checksum
3B219271042D6B13FCF5118C14EFFEB2

About this manual

The document describes a method using the Agilent 7700x ICP-MS equipped with the Octopole Reaction System (ORS) operating in Helium mode. This configuration allows for the simultaneous determination of trace and major elements in various food types while effectively suppressing interferences generated by the plasma and sample matrix.

To further increase throughput, an Agilent ISIS-DS discrete sampling system was evaluated. The study demonstrates that this combination enables a single analytical run to measure a wide range of elements across varying concentration levels with excellent agreement against certified reference materials.

What's inside

  • Introduction
  • Standard and sample preparation
  • Instrumentation
  • ISIS-DS discrete sampling system
  • Calibration
  • Interference suppression
  • Method validation

Frequently asked questions

How much sample and acid are required for microwave digestion?

The procedure requires 250 mg of sample digested with 3 mL of nitric acid and 1 mL of hydrogen peroxide.

What is the total analysis time per sample with the discrete sampling system?

Using the ISIS-DS system reduces the total analysis time per sample to 1.2 minutes.

What is the maximum concentration of matrix in the prepared samples?

The resulting samples contained up to 5 g/L of matrix.