Agilent SurePrint G3 Gene Expression Arrays Application Note

Gene expression analysis is currently carried out using three methods that provide complementary data: microarrays, RNA-Seq, and RT-qPCR. Microarrays provide comprehensive whole-genome coverage in single experiments, with rapid data acquisition and results at a low price per sample.


Agilent SurePrint G3 Gene Expression Arrays Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
416 KB
Published
08 June, 2012
Updated
30 September, 2023
MD5 checksum
34E6223DD42F6709C1ACA327CE44AC47

About this manual

This document describes a study validating the reliability of Agilent SurePrint G3 Gene Expression Microarrays. The authors compare microarray data against two complementary techniques: reverse transcriptase quantitative polymerase chain reaction (RT-qPCR) and next-generation sequencing (RNA-Seq). The goal is to demonstrate that the microarray platform provides accurate, reproducible results consistent with other established methods.

The study utilized specific Agilent reagents and systems, including the Low Input Quick Amp Labeling Kit for sample preparation and the Mx3005P QPCR System for RT-qPCR analysis. Results showed a correlation of R > 0.99 between microarray signals across technical replicates and a high concordance (R > 0.99) between microarray ratios and RT-qPCR predictions, confirming the accuracy of the SurePrint G3 arrays.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Microarray Analysis
  • RT-gPCR Analysis
  • RNA-Seq Analysis
  • Results and Discussion

Frequently asked questions

What is the minimum amount of total RNA required for labeling SurePrint G3 microarrays?

Twenty-five ng of total RNA was amplified and labeled using the Agilent Low Input Quick Amp Labeling Kit.

What is the minimum correlation coefficient (R) between SurePrint G3 microarray signals across technical replicates?

The reproducibility demonstrated a correlation of R > 0.99 for signals across technical replicates in one-color assays.

What is the minimum length of the single-end sequencing reads used in the RNA-Seq comparison?

The single end sequencing was performed with short 36 bp reads.