Agilent RapidFire High-Throughput Mass Spectrometry System Application Note

This application note describes an ultrafast SPE/MS/MS method, using the Agilent RapidFire High-Throughput Mass Spectrometry System, with a sample analysis time of <10 seconds per sample, for the analysis of levetiracetam in human serum. The Agilent RapidFire/MS System exhibited comparable accuracy, precision, linearity, and sensitivity to LC/MS/MS, but with a sample analysis cycle time 10-fold faster than LC/MS/MS.


Agilent RapidFire High-Throughput Mass Spectrometry System Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
568 KB
Published
11 June, 2012
Updated
01 October, 2023
MD5 checksum
B384055F2CB57027EC4DD93FF5589050

About this manual

This document evaluates the Agilent RapidFire High-Throughput Mass Spectrometry System for analyzing levetiracetam in human serum. The study demonstrates that the system achieves analysis times of less than 10 seconds per sample, offering a 10-fold speed increase over traditional LC/MS/MS methods while maintaining comparable accuracy, precision, and sensitivity.

The RapidFire/MS/MS system consists of an Agilent RapidFire 360, an Agilent 6460 Triple Quadrupole LC/MS System, and specific software. The document details the experimental conditions, sample preparation procedures, and results showing excellent linearity and low carryover, confirming the system's suitability for high-throughput analysis of similar assays.

What's inside

  • Abstract
  • Experimental
  • Sample preparation
  • RapidFire conditions
  • Triple quadrupole conditions
  • Results and Discussion
  • Conclusions

Specifications

Analysis time per sample< 10 seconds
Sample analysis cycle time9.5 seconds
Linear range1 pg/mL to 100 pg/mL
Correlation coefficient (R²)> 0.995
Signal-to-noise ratio> 20:1
Carryover< 1%

Frequently asked questions

How fast is the analysis cycle time for the Agilent RapidFire system?

The system is capable of analysis times of less than 10 seconds per sample.

Does the RapidFire system provide comparable results to LC/MS/MS?

Yes, the system had comparable accuracy, precision, linearity, and sensitivity to LC/MS/MS.

Download PDF (568 KB)