Hanna Instruments HI 98121 Water Ionizer pH/ORP Tester Supplemental Instruction Manual

Supplemental instructions for the Hanna Instruments HI 98121 pH/ORP tester in water ionizer applications, covering cleaning, ORP pretreatment and storage.


Hanna Instruments HI 98121 Water Ionizer pH/ORP Tester Supplemental Instruction Manual - cover page
Brand
Hanna Instruments
Category
Analyzer
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
428 KB
Published
09 November, 2026
Updated
09 November, 2026
MD5 checksum
11F9FE41ED9BF87FA33AD9F6200A8E92

About this manual

This supplemental instructional manual from Hanna Instruments covers use of the HI 98121 portable pH and ORP tester for water ionizer applications. It answers frequent questions about why the tester may stop reading high negative mV ORP values and how often the pH/ORP sensor should be cleaned in HI 7061 cleaning solution.

The manual explains the pH scale from 0 to 14 and the meaning of ORP in millivolts, then describes ORP preparation and measurement, including pretreatment with HI 7091M or HI 7092M solutions and a pH/mV reference table. Maintenance, cleaning, polishing of the platinum sensor and storage with HI 70300M storage solution are also described, along with the solutions contained in the ionizer solutions kit.

What's inside

  • Frequent Questions and Answers
  • pH Scale
  • Background
  • Ionizer Solutions Kit
  • ORP Preparation and Measurement
  • Maintenance & Cleaning
  • Storage

Specifications

Measured parameterspH and ORP
Form factorPortable tester

Frequently asked questions

Why doesn't my tester read high -mV ORP (-250mV to -999mV) anymore?

Mineral buildup on the surface of the platinum ORP sensor interferes with the full -mV scale. Try cleaning the sensor in HI7061 cleaning solution for 15 minutes.

How often should I clean my pH/ORP sensor in HI7061 cleaning solution?

Clean the pH/ORP sensor after each use by soaking it in HI7061 cleaning solution for 2-3 minutes.

How long should I pre-treat the ORP platinum surface before use?

Soak the platinum surface in pretreatment solution for 15 minutes to get a faster response time.

How should the electrode be stored?

Keep the tip and junction moist by replacing the protective cap with a few drops of HI 70300M storage solution, and never store the electrode in distilled or deionized water.