Endevco 7264C Manuals

ManualLib has 16 documents for the Endevco 7264C (sensor): 16 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 16

Key facts

Brand
Endevco
Model
7264C
Category
sensor
Industry
Industrial
Documents
16
Languages
EN
Last added
2023-10-18

Start here

Endevco 7264C Piezoresistive Accelerometer Instruction Manual

Endevco 7264C Piezoresistive Accelerometer Instruction Manual

User Manual 3 pages · 226 KB · EN

This manual provides instructions for the Endevco Model 7264C, a low mass, miniature piezoresistive accelerometer. It details necessary precautions to avoid damage during handling and mounting, including torque specifications and the use of supplied materials like Allen wrenches and washers.

The document outlines a check-out procedure for verifying proper operation upon receipt, including impedance, insulation resistance, and zero measurand output tests. It also covers mounting techniques, cable management, and guidelines for recalibration every 6 to 12 months.

Read online

Endevco 7264C specifications

TypePiezoresistive Accelerometer
Recommended Mounting Torque2.6 in-lbf (0.29 N-m)
Maximum Mounting Torque3.0 in-lbf (0.34 N-m)
Minimum Insulation Resistance100 megohms
Recommended Surface Roughness32 microinch rms

Source: Endevco 7264C Piezoresistive Accelerometer Instruction Manual

All Endevco 7264C documents by type

Questions answered in these manuals

How often should the accelerometer be recalibrated?

Sensitivity and ZMO calibration should be performed at 6 to 12 month intervals, depending on usage.

From: Endevco 7264C Piezoresistive Accelerometer Instruction Manual

What is the maximum torque allowed for mounting?

The maximum recommended mounting torque is 3.0 in-lbf (0.34 N-m).

From: Endevco 7264C Piezoresistive Accelerometer Instruction Manual

What is the recommended surface roughness for mounting?

The recommended surface roughness is 32 microinch rms.

From: IM7264C accelerometer instruction manual

How often should recalibration be performed?

Sensitivity and ZMO calibration should be performed at 6 to 12 month intervals.

From: IM7264C accelerometer instruction manual

What is the maximum recommended mounting torque for the 7264C?

The maximum recommended mounting torque is 3.0 in-lbf (0.34 N-m).

From: IM7264C instruction manual

What is the minimum acceptable insulation resistance?

The insulation resistance between leads/shield and case must be at least 100 megohms.

From: IM7264C instruction manual