IM7264C is the instruction manual for Endevco's 7264C sensor
This manual provides instructions for installing, testing, and maintaining the Endevco 7264C piezoresistive accelerometer. It includes precautions for handling, mounting torque specifications, and procedures for impedance and insulation resistance tests. The document also covers recalibration intervals and cleaning guidelines.
The manual specifies required mounting materials, torque limits, and cable handling to prevent damage. It includes a diagram for installation and contact information for technical support.
| Mounting Torque | 2.6 in-Ibf (0.29 N-m) recommended, 3.0 in-Ibf (0.34 N-m) maximum |
|---|---|
| Insulation Resistance | 100 megohms minimum |
| Zero Measurand Output | Within data sheet specified limits |
The recommended mounting torque is 2.6 in-Ibf (0.29 N-m), with a maximum of 3.0 in-Ibf (0.34 N-m).
Use a multimeter, ohmmeter, or megohmmeter set at 50 volts maximum to measure insulation resistance between all leads and the cable shield, transducer case, and cable shield to transducer case.
Sensitivity and ZMO calibration should be performed every 6 to 12 months, depending on usage.