Endevco Model 7264C Piezoresistive Accelerometer Instruction Manual

The Model 7264C is a low mass, miniature piezoresistive accelerometer produced by Endevco.


Endevco Model 7264C Piezoresistive Accelerometer Instruction Manual - cover page
Brand
Endevco
Category
Other Electronics
Document type
User Manual
Language
English
Pages
3
File format
PDF
File size
226 KB
Published
18 October, 2023
Updated
18 October, 2023
Model
7264C
MD5 checksum
E22953989D02BA34D3E0ABE7899FFF65

About this manual

The Endevco Model 7264C is a low mass, miniature piezoresistive accelerometer. The manual provides specific precautions for handling and mounting the unit to avoid damage, including warnings against dropping the unit and over-torquing the mounting screws. It lists the materials supplied with the unit, such as Allen wrenches and socket head cap screws, and details the correct installation torque of 2.6 in-lbf (0.29 N-m) with a maximum of 3.0 in-lbf (0.34 N-m).

The document outlines a check-out procedure for verifying proper operation upon receipt, including impedance tests, insulation resistance checks, and zero measurand output verification. It also describes the mounting process, emphasizing the need for a clean surface and proper cable management, and provides guidelines for recalibration intervals and cleaning the unit.

What's inside

  • Materials Required
  • Special Precautions
  • Check-out
  • Zero measurand Output
  • Mounting
  • Recalibration
  • Cleaning

Specifications

Product Typelow mass, miniature piezoresistive accelerometer
Recommended Mounting Torque2.6 in-lbf (0.29 N-m)
Maximum Mounting Torque3.0 in-lbf (0.34 N-m)
Recommended Surface Roughness32 microinch rms
Minimum Insulation Resistance100 megohms

Frequently asked questions

What is the recommended mounting torque for the Model 7264C?

The recommended mounting torque is 2.6 in-lbf (0.29 N-m), with a maximum of 3.0 in-lbf (0.34 N-m).

How often should the accelerometer be recalibrated?

Sensitivity and ZMO calibration should be performed at 6 to 12 month intervals, depending on usage.

How should the unit be cleaned?

Dirty units may be wiped clean using a damp cloth and a solvent such as acetone, but should not be soaked or immersed.