IM7264C instruction manual
This manual provides instructions for handling, mounting, and testing the Endevco 7264C piezoresistive accelerometer. It includes precautions to avoid damage, such as avoiding drops and using correct torque. Testing procedures cover impedance, insulation resistance, and zero measurand output. Mounting requires specific tools and techniques, including using supplied washers and screws. Recalibration is recommended every 6 to 12 months depending on usage.
The recommended mounting torque is 2.6 in-lbf (0.29 N-m), with a maximum of 3.0 in-lbf (0.34 N-m).
Dirty units may be wiped clean using a damp cloth and a solvent such as acetone. DO NOT SOAK OR IMMERSE the unit in any solvent or water.