Endevco Model 7264C Piezoresistive Accelerometer Instruction Manual

IM7264C model low mass miniature piezoresistive accelerometer instruction manual, introduces the use of accelerometer, installation method and check method.


Endevco Model 7264C Piezoresistive Accelerometer Instruction Manual - cover page
Brand
Endevco
Category
sensor
Document type
User Manual
Language
English
Pages
3
File format
PDF
File size
226 KB
Published
18 October, 2023
Updated
18 October, 2023
Model
7264C
MD5 checksum
86ADE4C841E40F62034460E8BB880A61

About this manual

The Endevco Model 7264C is a low mass, miniature piezoresistive accelerometer. The document outlines specific precautions for handling and mounting to avoid damage, including avoiding drops and controlling torque. It lists the materials supplied with the unit, such as Allen wrenches and screws.

What's inside

  • Materials Required
  • Special Precautions
  • Check-out
  • Mounting
  • Installation
  • Recalibration
  • Cleaning

Specifications

Mounting Torque (Recommended)2.6 in-lbf (0.29 N-m)
Mounting Torque (Maximum)3.0 in-lbf (0.34 N-m)
Impedance Tolerance±25% of calibrated value
Insulation Resistance Minimum100 megohms
Recommended Surface Roughness32 microinch rms

Frequently asked questions

What is the recommended mounting torque for the accelerometer?

The recommended mounting torque is 2.6 in-lbf (0.29 N-m).

How often should the accelerometer be recalibrated?

Sensitivity and ZMO calibration should be performed at 6 to 12 month intervals.

How should the unit be cleaned?

Dirty units may be wiped clean using a damp cloth and a solvent such as acetone.