Endevco IM7264B Manuals

ManualLib has 6 documents for the Endevco IM7264B: 6 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 6

Key facts

Brand
Endevco
Model
IM7264B
Documents
6
Languages
EN
Last added
2023-10-18

Start here

Endevco IM7264B Accelerometer Instruction Manual

Endevco IM7264B Accelerometer Instruction Manual

User Manual 3 pages · 157 KB · EN

The Endevco IM7264B is a low mass, miniature piezoresistive accelerometer. Due to its small size, specific precautions must be observed during handling and mounting to avoid damaging the unit. The document lists the materials supplied with the unit, including Allen wrenches, washers, and screws, and details special precautions such as avoiding drops and preventing over-torquing.

Read online

Endevco IM7264B specifications

Mounting Torque (Recommended)2.6 in-lbf (0.29 N-m)
Mounting Torque (Maximum)3.0 in-lbf (0.34 N-m)
Minimum Insulation Resistance100 megohms
Impedance Tolerance±25% of calibrated value

Source: Endevco IM7264B Accelerometer Instruction Manual

All Endevco IM7264B documents by type

Questions answered in these manuals

How often should the accelerometer be recalibrated?

Sensitivity and ZMO calibration should be performed at 6 to 12 month intervals, depending on usage.

From: Endevco IM7264B Accelerometer Instruction Manual

How should the accelerometer be cleaned?

Dirty units may be wiped clean using a damp cloth and a solvent such as acetone, but must not be soaked or immersed.

From: Endevco IM7264B Accelerometer Instruction Manual

What is the recommended mounting torque for the IM7264B?

The recommended mounting torque is 2.6 in-lbf (0.29 N-m), with a maximum of 3.0 in-lbf (0.34 N-m).

From: Endevco IM7264B Accelerometer Instruction Manual

What is the minimum insulation resistance for the IM7264B?

The insulation resistance between leads/shield and case should be at least 100 megohms.

From: Endevco IM7264B Accelerometer Instruction Manual

What is the recommended mounting torque for the IM7264B?

The recommended mounting torque is 2.6 in-lbf (0.29 N-m), with a maximum of 3.0 in-lbf (0.34 N-m).

From: Endevco IM7264B Piezoresistive Accelerometer Instruction Manual

How should the IM7264B be cleaned?

Dirty units may be wiped clean using a damp cloth and a solvent such as acetone. DO NOT SOAK OR IMMERSE the unit in any solvent or water.