Endevco IM7264B Accelerometer Instruction Manual
The Endevco IM7264B is a low mass, miniature piezoresistive accelerometer. Due to its small size, specific precautions must be observed during handling and mounting to avoid damaging the unit. The document lists the materials supplied with the unit, including Allen wrenches, washers, and screws, and details special precautions such as avoiding drops and preventing over-torquing.
Read onlineEndevco IM7264B specifications
| Mounting Torque (Recommended) | 2.6 in-lbf (0.29 N-m) |
|---|---|
| Mounting Torque (Maximum) | 3.0 in-lbf (0.34 N-m) |
| Minimum Insulation Resistance | 100 megohms |
| Impedance Tolerance | ±25% of calibrated value |
All Endevco IM7264B documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
- Endevco IM7264B Accelerometer Instruction Manual 3 p
- Endevco IM7264B Accelerometer Instruction Manual 3 p
- Endevco IM7264B Piezoresistive Accelerometer Instruction Manual 3 p
- Endevco IM7264B Piezoresistive Accelerometer Instruction Manual 3 p
- IM7264B piezoresistive accelerometer installation instructions 3 p
- IM7264B Piezoresistive Accelerometer instruction manual 3 p
Questions answered in these manuals
How often should the accelerometer be recalibrated?
Sensitivity and ZMO calibration should be performed at 6 to 12 month intervals, depending on usage.
How should the accelerometer be cleaned?
Dirty units may be wiped clean using a damp cloth and a solvent such as acetone, but must not be soaked or immersed.
What is the recommended mounting torque for the IM7264B?
The recommended mounting torque is 2.6 in-lbf (0.29 N-m), with a maximum of 3.0 in-lbf (0.34 N-m).
What is the minimum insulation resistance for the IM7264B?
The insulation resistance between leads/shield and case should be at least 100 megohms.
What is the recommended mounting torque for the IM7264B?
The recommended mounting torque is 2.6 in-lbf (0.29 N-m), with a maximum of 3.0 in-lbf (0.34 N-m).
From: Endevco IM7264B Piezoresistive Accelerometer Instruction Manual
How should the IM7264B be cleaned?
Dirty units may be wiped clean using a damp cloth and a solvent such as acetone. DO NOT SOAK OR IMMERSE the unit in any solvent or water.
From: Endevco IM7264B Piezoresistive Accelerometer Instruction Manual