User Manual 3 pages · 122 KB · EN

This manual provides instructions for installing, handling, and testing the Endevco 7264D piezoresistive accelerometer. It includes precautions to avoid damage, torque specifications, and procedures for impedance, insulation resistance, and zero measurand output tests. Recalibration is recommended every 6 to 12 months depending on usage. Cleaning instructions and contact information for technical support are also provided.

Read online

Endevco 7264D specifications

TypePiezoresistive Accelerometer
Mounting Torque (Recommended)2.6 in-lbf (0.29 N-m)
Mounting Torque (Maximum)3.0 in-lbf (0.34 N-m)
Surface Roughness32 microinch rms
Insulation Resistance100 megohms minimum

Source: Endevco Model 7264D Piezoresistive Accelerometer Instruction Manual

All Endevco 7264D documents by type

Questions answered in these manuals

What is the maximum recommended mounting torque for the Model 7264D accelerometer?

The maximum specified mounting torque is 3.0 in-lbf (0.34 N-m).

From: Endevco 7264D Piezoresistive Accelerometer Instruction Manual

What is the minimum required insulation resistance for the accelerometer leads?

All insulation resistance readings must be at least 100 megohms.

From: Endevco 7264D Piezoresistive Accelerometer Instruction Manual

What is the recommended mounting torque for the 7264D accelerometer?

The specified mounting torque is 2.6 in-lbf (0.29 N-m) recommended, 3.0 in-lbf (0.34 N-m) maximum.

From: Endevco 7264D Piezoresistive Accelerometer Instruction Manual

How should the accelerometer be cleaned?

Dirty units may be wiped clean using a damp cloth and a solvent such as acetone. DO NOT SOAK OR IMMERSE the unit in any solvent or water.

From: Endevco 7264D Piezoresistive Accelerometer Instruction Manual

How often should the accelerometer be recalibrated?

Recalibration should be performed at 6 to 12 month intervals, depending on usage.

From: Endevco Model 7264D Piezoresistive Accelerometer Instruction Manual

What is the minimum insulation resistance reading?

All readings for insulation resistance should be 100 megohms minimum.

From: Endevco Model 7264D Piezoresistive Accelerometer Instruction Manual