Endevco 7264B Piezoresistive Accelerometer User Manual
This manual provides instructions for mounting, handling, and checking out the Endevco 7264B piezoresistive accelerometer. It details required materials, torque specifications, and initial tests to verify proper operation. The document also covers installation best practices, cable handling, and recalibration intervals. Cleaning instructions and contact information for support are included.
Read onlineEndevco 7264B specifications
| Mounting Torque (Recommended) | 2.6 in-lbf (0.29 N-m) |
|---|---|
| Mounting Torque (Maximum) | 3.0 in-lbf (0.34 N-m) |
| Minimum Insulation Resistance | 100 megohms |
| Impedance Tolerance | ±25% of calibrated value |
| Recommended Surface Roughness | 32 micro inch rms |
Source: Endevco Model 7264B Piezoresistive Accelerometer Instruction Manual
All Endevco 7264B documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
- Endevco 7264B Piezoresistive Accelerometer User Manual 3 p
- IM7264B instruction manual 3 p
- Endevco 7264B Piezoresistive Accelerometer Instruction Manual 3 p
- Endevco Model 7264B Piezoresistive Accelerometer Instruction Manual 3 p
- Endevco 7264B Piezoresistive Accelerometer Installation Manual 3 p
- Instruction manual for IM7264B piezoelectric accelerometer 3 p
- IM7264B model accelerometer instruction manual 3 p
- Endevco Model 7264B Piezoresistive Accelerometer Instruction Manual 3 p
- Endevco 7264B Piezoresistive Accelerometer Instruction Manual 3 p
- IM7264B Instruction Manual 3 p
- Endevco 7264B Piezoresistive Accelerometer Instruction Manual 3 p
- IM7264B piezoresistive accelerometer installation instructions 3 p
- IM7264B Piezoresistive Accelerometer instruction manual 3 p
Questions answered in these manuals
What is the recommended mounting torque for the 7264B?
The recommended mounting torque is 2.6 in-lbf (0.29 N-m), with a maximum of 3.0 in-lbf (0.34 N-m).
From: Endevco 7264B Piezoresistive Accelerometer User Manual
How should I test the accelerometer upon receipt?
Perform three tests: impedance test, insulation resistance test, and zero measurand output test using specified equipment and procedures.
From: Endevco 7264B Piezoresistive Accelerometer User Manual
What is the recommended mounting torque for the 7264B accelerometer?
The recommended mounting torque is 2.6 in-lbf (0.29 N-m), with a maximum of 3.0 in-lbf (0.34 N-m).
How should the accelerometer be cleaned?
Dirty units may be wiped clean using a damp cloth and a solvent such as acetone. DO NOT SOAK OR IMMERSE the unit in any solvent or water.
What is the maximum recommended mounting torque for the accelerometer?
The maximum recommended mounting torque is 3.0 in-lbf (0.34 N-m).
From: Endevco 7264B Piezoresistive Accelerometer Instruction Manual
What is the minimum required insulation resistance reading?
All insulation resistance readings must be at least 100 megohms.