IM7264C Sensor User Manual

This document introduces the use of IM7264C sensors, including material requirements, special precautions, and inspection methods.


IM7264C Sensor User Manual - cover page
Brand
Endevco
Category
sensor
Document type
User Manual
Language
English
Pages
3
File format
PDF
File size
226 KB
Published
18 October, 2023
Updated
18 October, 2023
Model
7264C
MD5 checksum
70A0174475F8ABDB15CF37E45F109783

About this manual

This document is the instruction manual for the Endevco Model 7264C, a low mass, miniature piezoresistive accelerometer. It outlines specific precautions for handling and mounting the unit to prevent damage, including torque limits and the use of supplied materials like Allen wrenches and washers.

The guide provides step-by-step instructions for initial check-out procedures, such as impedance and insulation resistance tests, as well as zero measurand output verification. It details the correct mounting techniques, including surface preparation and cable management, and specifies a recalibration schedule of 6 to 12 months depending on usage.

What's inside

  • Materials Required
  • Special Precautions
  • Check-out
  • Mounting
  • Installation
  • Recalibration
  • Cleaning

Specifications

TypePiezoresistive Accelerometer
Recommended Mounting Torque2.6 in-lbf (0.29 N-m)
Maximum Mounting Torque3.0 in-lbf (0.34 N-m)
Minimum Insulation Resistance100 megohms
Recommended Surface Roughness32 microinch rms

Frequently asked questions

How often should the accelerometer be recalibrated?

Sensitivity and ZMO calibration should be performed at 6 to 12 month intervals, depending on usage.

What is the minimum insulation resistance for the unit?

All readings for insulation resistance should be a minimum of 100 megohms.

Can the unit be cleaned with solvents?

Dirty units may be wiped clean with a damp cloth and a solvent such as acetone, but the unit must not be soaked or immersed.