This document is the user manual for RIGOL RP1003C/RP1004C/RP1005C current probes, which introduces the product's warranty and declaration, trademark information, product certification, general safety overview, safety terms and symbols, etc.
The document serves as a comprehensive guide for the RIGOL RP1003C, RP1004C, and RP1005C Current Probes. It details general safety precautions, including grounding requirements and environmental conditions to prevent damage or injury during operation.
The manual provides step-by-step instructions for measurement preparation, demagnetization, zero adjustment, and the actual measurement process. It also includes specific technical specifications for each model variant regarding bandwidth, input current range, gain, and physical dimensions. This manual is written in Chinese.
| Bandwidth (RP1003C) | DC to 50MHz (-3dB) |
|---|---|
| Rise Time | ≤7ns |
| Input Current Linear Response Range (RP1003C) | 30Arms |
| Max Peak Current (RP1003C) | 50A peak, non-continuous |
| Gain | 0.1V/A |
| Power Supply Voltage | +12V±0.5 V |
| Max Rated Voltage (RP1003C/RP1004C) | 300V, CAT I |
| Max Rated Voltage (RP1005C) | 4000V transient voltage |
| Weight (RP1003C) | approx. 230g |
The RP1005C supports Measurement Category II and III.
The RP1003C has a maximum peak current of 50A, which is rated for non-continuous use.
The RP1005C current sensor measures approximately 176mm x 69mm x 34mm.
The RP1005C can measure conductors with a maximum diameter of approximately 20mm.