Agilent Method development on the Agilent 1290 Infinity LC using Intelligent System Emulation Technology (ISET) with subsequent transfer to an Agilent 1100 Series LC Manual

This Technical Overview shows the development of a chromatographic method for the analysis of paracetamol and its impurities using 1290 Infi nity LC System with ISET. Having developed the method it was transferred to an Agilent 1100 Series Quaternary LC System.


Agilent Method development on the Agilent 1290 Infinity LC using Intelligent System Emulation Technology (ISET) with subsequent transfer to an Agilent 1100 Series LC Manual - cover page
Brand
Agilent
Document type
Other
Language
English
Pages
8
File format
PDF
File size
846 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
4CDC924BC9A9B23D4EE0BA17CBC68073

About this manual

Agilent's Intelligent System Emulation Technology (ISET) allows for the seamless transfer of methods between different Liquid Chromatography (LC) systems. This document demonstrates how a method developed on the high-performance Agilent 1290 Infinity LC System can be emulated to predict performance on an older Agilent 1100 Series Quaternary LC System before physical transfer.

What's inside

  • Introduction
  • Instrumentation and software
  • Sample
  • Chromatographic conditions
  • Method development on the 1290 Infinity LC System
  • Emulation of the 1100 Series Quaternary LC System using ISET
  • Results and discussion
  • Conclusion

Specifications

ColumnAgilent ZORBAX SB C18, 150 x 4.6 mm, 5 µm
Mobile phase AWater + 0.1% TFA
Mobile phase BAcetonitrile + 0.09% TFA
Flow rate1.0 mL/min
Detection wavelengths220, 254, 270, 310/10 nm

Frequently asked questions

What is the retention time deviation when using ISET compared to the 1100 Series?

The retention time deviation for the 1290 Infinity LC System with ISET was between 0.5% and 1.9%.

What is the resolution deviation when using ISET compared to the 1100 Series?

The deviation of the resolution on the 1290 Infinity LC System with ISET was between 0.11% and 2.4%.

What software revision was used for the experiments?

Agilent ChemStation revision C.01.03 and ISET revision 1.0 were used.