Agilent 1260 Infinity Bio-inert Quaternary LC System Technical Overview

The Agilent 1260 Infinity Bio-inert Quaternary LC system has the same performance specifications as the standard Agilent 1260 Infinity Quaternary LC system based on the proven technology of the Agilent 1200 Infinity Series. This Technical Overview proves that method transfer from an Agilent 1260 Infinity Series Quaternary LC system to an Agilent 1260 Infinity Bio-inert Quaternary LC system will typically result in chromatograms with no significant differences including compatibility with standard methods with a maximum pressure of 600 bar. In addition, it is especially well suited for bio-analytical applications, such as, protein analysis using ion exchange chromatography (IEX) or size exclusion/gel filtration (SEC).


Agilent 1260 Infinity Bio-inert Quaternary LC System Technical Overview - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
12
File format
PDF
File size
1.3 MB
Published
05 July, 2012
Updated
29 September, 2023
MD5 checksum
046B4AD18DCF69AC5AB3C5DB336FAD5C

About this manual

This document serves as a technical overview for the Agilent 1260 Infinity Bio-inert Quaternary LC system, designed specifically for bio-chromatography applications. It demonstrates that the system maintains the same performance specifications as the standard 1260 Infinity Quaternary LC system, supporting a maximum pressure of 600 bar and compatibility with standard methods.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Pump performance
  • Injector performance
  • Low back pressure application

Specifications

Maximum pressure600 bar
Maximum temperature80 °C
Flow rate rangeup to 5 mL/min
Retention time precision (gradient)< 0.08 % RSD
Retention time precision (fast gradient)< 0.11% RSD
Retention time precision (isocratic)< 0.1% RSD
Ripple on 10% step0.07%
Ripple on 50% step0.06%
Ripple on 90% step0.07%
System delay volume1,000 pL
Area precision (1 pL injection)<0.33 %
Area precision (0.5 pL injection)<0.75%

Frequently asked questions

Is the 1260 Infinity Bio-inert Quaternary LC system compatible with standard methods?

Yes, it is compatible with standard methods with a maximum pressure of 600 bar.

What is the maximum pressure supported by the system?

The system supports a maximum pressure of 600 bar.

What is the maximum temperature capability of the system?

The system has a maximum temperature of 80 °C, or 100 °C with an Agilent 1290 Infinity Thermostatted Column Compartment.

Is the system suitable for protein analysis?

Yes, it is especially well suited for bio-analytical applications, such as protein analysis using ion exchange chromatography or size exclusion/gel filtration.