Agilent Substituted Phenols Detector Response Application Note

The document highlights the advantages of using ZORBAX SB-C18 columns for gradient separations of complex environmental mixtures, demonstrating resolution >1.5 and good peak shapes for all components. The conditions used by Robert Ricker in the environmental analysis application are ZORBAX SB-C18 column, 20% ACN/80% 0.01 M H3PO4 solution, gradient elution from 20% ACN to 45% ACN in 7.5 minutes, followed by gradient elution from 45% ACN to 80% ACN in 2.0 minutes, flow rate of 1.5 mL/min, and temperature of 35°C.


Agilent Substituted Phenols Detector Response Application Note - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
946 KB
Published
03 July, 2012
Updated
29 September, 2023
MD5 checksum
6E90E72B3A7E1FD0FA485391332659A8

About this manual

This document presents the detector response for a mixture of substituted phenols, including environmental samples. It demonstrates the chromatographic performance of Agilent ZORBAX SB-C18 columns in separating these complex mixtures.

The application highlights that gradient separations are well-suited for analyzing complex environmental mixtures using this column technology.

What's inside

  • Detector Response
  • Application
  • Environmental Analysis
  • Substituted Phenols List
  • Chromatographic Conditions

Frequently asked questions

What column is used for the gradient separations of complex environmental mixtures?

The ZORBAX SB-C18 column, with part number 883975-902, is used for these separations.

What is the flow rate and temperature for the chromatographic conditions?

The flow rate is 1.5 mL/min and the temperature is set to 35°C.

What is the resolution achieved for all components in the separation?

All components have a resolution greater than 1.5 with good peak shapes.