Agilent Biogenic Amines Catecholamines Application Note

The document describes a method for high-performance liquid chromatography (HPLC) analysis using ion-pair reversed-phase separations combined with electrochemical detection to measure the levels of catecholamines and indoleamines in brain tissue. The method is rapid, selective, and efficient, allowing for analysis of samples in less than 10 minutes per sample. The use of a highly stable column-packing material (Agilent ZORBAX SB-C8, 3.5 µm) ensures unmatched column lifetime, enabling analysis of thousands of samples per column. Samples are directly injected as perchloric acid (0.1N) extracts, eliminating the need for sample workup. This method is suitable for the determination of various biogenic amines neurotransmitters and their metabolites.


Agilent Biogenic Amines Catecholamines Application Note - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
2.1 MB
Published
03 July, 2012
Updated
29 September, 2023
MD5 checksum
753A7E0F3146D19E94B6F19F17BF1F2A

About this manual

This document details a High-Performance Liquid Chromatography (HPLC) method for analyzing neurochemicals such as dopamine, norepinephrine, and epinephrine in mouse brain homogenates. The procedure utilizes ion-pair reversed-phase separations combined with electrochemical detection to achieve rapid and selective analysis.

What's inside

  • Application
  • Highlights
  • Neurochemicals Analysis
  • Column Packing Material
  • Sample Preparation
  • Chromatographic Conditions

Frequently asked questions

How long does the analysis of neurochemicals in tissue homogenates take?

The rapid and selective analysis is completed in less than 10 minutes per sample.

What type of column material is used for the separation shown?

The separation uses a smaller particle, highly stable, column-packing material (Agilent ZORBAX SB-C8).

What is the flow rate for the HPLC analysis conditions?

The mobile phase flows at 1.5 mL / min.

Download PDF (2.1 MB)

More Agilent Manuals