Agilent ZORBAX SB-C8 Application Note Antihistamines

This document describes the features of ZORBAX StableBond C8, which has a sterically protected bonded phase that allows stable operation at low mobile-phase pH. ZORBAX SB-C8 separates basic compounds with good resolution and peak shape using simple mobile phases. The document also provides the conditions for using ZORBAX Rx/SB-C8 in pharmaceutical applications.


Agilent ZORBAX SB-C8 Application Note Antihistamines - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
354 KB
Published
03 July, 2012
Updated
29 September, 2023
MD5 checksum
702B5D3EDA94B5C761355B797854D12F

About this manual

This document presents an application demonstrating the separation of four antihistamines: Pseudoephedrine, Scopolamine, Pyrilamine, and Chlorpheniramine. The analysis utilizes the ZORBAX StableBond C8 column, which features a sterically protected bonded phase allowing for stable operation at low mobile-phase pH.

The chromatographic method employs a simple mobile phase consisting of 10% acetonitrile and 90% 50mM KH2PO4 at pH 2.5. The separation is performed on a 4.6 x 150 mm column at 25°C with a flow rate of 2.0 mL/min and UV detection at 254 nm.

What's inside

  • Application
  • Highlights
  • Chromatogram
  • Conditions
  • Copyright

Specifications

Column Dimensions4.6 x 150 mm
Mobile Phase10% ACN, 90% 50mM KH2PO4, pH 2.5
Flow Rate2.0 mL/min
Detection Wavelength254 nm
Part Number883967-901 / 883975-906

Frequently asked questions

Which column is used for separating antihistamines?

The ZORBAX StableBond C8 column is used for the separation.

What is the mobile phase composition?

The mobile phase consists of 10% acetonitrile and 90% 50mM KH2PO4 at pH 2.5.