Agilent Technologies 5988-6440EN Application Note Use of Column Configuration to Increase Resolution and Decrease Run Ti

The use of smaller particle-size columns can increase resolution and peak height, while decreasing run time.


Agilent Technologies 5988-6440EN Application Note Use of Column Configuration to Increase Resolution and Decrease Run Ti - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
940 KB
Published
03 July, 2012
Updated
29 September, 2023
MD5 checksum
6C07C242FCFE2B3348C869322506713A

About this manual

This document presents an application note by Robert Ricker regarding chromatographic separation techniques. It highlights how transferring separations to columns with smaller particle sizes (specifically 3.5 µm) can result in increased resolution and peak height compared to larger particle sizes (5 µm).

What's inside

  • Application
  • Highlights
  • Conditions
  • Copyright

Specifications

Column TypeZORBAX 300SB-C8 Columns
Mobile Phase A95% ACN : 5% H2O, 0.1% TFA
Mobile Phase B5% ACN : 95% H2O, 0.085% TFA
Injection Volume10pL
Flow Rate1 mL/min
Column Temperature35°C
Detection MethodUV (245 nm)

Frequently asked questions

What are the dimensions and particle size of the ZORBAX 300SB-C8 columns used in this application?

The document lists three column sizes: 4.6 x 50 mm, 4.6 x 150 mm, and 4.6 x 250 mm, all with a particle size of either 3.5 um or 5 um.

What are the specific mobile phase conditions for the chromatographic separation described?

The method uses Mobile Phase A as 95% ACN with 5% H2O and 0.1% TFA, and Mobile Phase B as 5% ACN with 95% H2O and 0.085% TFA.

What are the injection volume, flow rate, and column temperature settings for this procedure?

The procedure requires an injection of 10 uL at a flow rate of 1 mL/min with the column maintained at 35°C.