Agilent 1100 Series HPLC Analysis of Doxepin in Aponal 100 Tablets Application Note

This document describes a new HPLC method for the determination of the tricyclic antidepressant doxepin in 'Aponal 100' tablets. The method uses an extra densely bonded Eclipse XDB-C18 column, which is made with a special Zorbax silica support and designed to reduce or eliminate strong adsorption of basic or acidic compounds. The method is stable at neutral pH and shows excellent and rapid resolution of tricyclic aromatic amines at low buffer concentration. Doxepin is separated from 'Aponal 100' tablets with no interference from excipients.


Agilent 1100 Series HPLC Analysis of Doxepin in Aponal 100 Tablets Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
38 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
D8AF9A2942EA3CBDFF2B465E8F4256A6

About this manual

This document details the high-performance liquid chromatography (HPLC) analysis of the tricyclic antidepressant doxepin in 'Aponal 100' tablets. The method utilizes an Agilent 1100 Series HPLC system equipped with a Zorbax Eclipse XDB-C18 column to achieve excellent peak shape and rapid resolution of tricyclic aromatic amines at neutral pH. The analysis successfully separates doxepin from excipients and other tricyclic antidepressants like imipramine and amitriptyline.

What's inside

  • Tricyclic Antidepressant
  • HPLC Analysis
  • Doxepin in Aponal 100 Tablets
  • Eclipse XDB-C18 Column
  • Experimental Conditions
  • Chromatogram Results
  • Agilent 1100 Series HPLC

Specifications

EquipmentAgilent 1100 Series HPLC
Mobile Phase A0.025 M KH2PO4 in water (pH = 7)
Mobile Phase BMethanol
Temperature25 °C
Flow Rate1.0 mL/min
UV Detector Wavelength210 nm

Frequently asked questions

Is doxepin stable under the analysis conditions?

Yes, doxepin is stable at neutral pH, which is the condition used in this method.

At what temperature should the autosampler be set?

The autosampler temperature must be set to 4 °C to avoid sample decomposition.

What is the flow rate used in this HPLC analysis?

The analysis is performed with a flow rate of 1.0 mL/min.