Agilent 1100 LC/MS Application Note for Sirolimus and Tacrolimus Analysis

This document describes an improved method for the simultaneous measurement of sirolimus and tacrolimus concentrations in blood, in which the drugs are first isolated from whole-blood specimens by solid phase extraction and then analyzed by liquid chromatography and electrospray-mass spectrometry.


Agilent 1100 LC/MS Application Note for Sirolimus and Tacrolimus Analysis - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
104 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
8EE63A99F8680F4070C54D637414465D

About this manual

This document details an improved analytical method for measuring the immunosuppressants sirolimus and tacrolimus in whole-blood specimens. The method utilizes solid phase extraction followed by analysis via liquid chromatography and electrospray ionization mass spectrometry (LC/ESI-MS). The procedure is designed to be semi-automated to increase sample throughput and productivity in clinical laboratories.

What's inside

  • Introduction
  • Chemical Structures
  • Method
  • LC Parameters
  • MS Parameters
  • Results
  • Precision
  • Limit of Detection
  • Interference Study
  • Method Comparison

Specifications

Limit of Detection (Sirolimus)0.38 ug/L
Analytical columnRestek Allure C18, 30 mm long x 3.2 mm id
Flow rate0.5 mL/min
Temperature75 °C

Frequently asked questions

Can sirolimus and tacrolimus be measured simultaneously?

Yes, both drugs are measured using the same extraction and LC/MS conditions.

What is the flow rate for the LC parameters?

The flow rate is set to 0.5 mL/min.