Agilent 6890 GC Combined GC/MSD and GC/FID Data Acquisition and Analysis Application Note

The document describes the ability to identify and quantify compounds based on single and dual column analyses. Instruments used include Agilent Technologies 6890 with other Hewlett-Packard instruments and data systems. The document demonstrates the acquisition and analysis of combined GC/MSD and GC/FID data.


Agilent 6890 GC Combined GC/MSD and GC/FID Data Acquisition and Analysis Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
418 KB
Published
02 July, 2012
Updated
29 September, 2023
MD5 checksum
66F0BDC498AFB298F47E4E213F5EB9B6

About this manual

This document describes a method for combining data from a Gas Chromatograph/Mass Spectrometer Detector (GC/MSD) and a Gas Chromatograph/Flame Ionization Detector (GC/FID) to increase the information available from analyses. The combined data allows for identification via mass spectra and quantitation via the GC detector, offering benefits such as faster method development and maximum information extraction. The note focuses on configurations using the Agilent 6890 Series GC and 5972 MSD controlled by the Productivity ChemStation.

What's inside

  • Introduction
  • Single Column Configuration
  • Dual Column Configuration
  • Signal Alignment
  • Experimental Conditions
  • MSD Parameters
  • FID Parameters
  • Column and Inlet Conditions

Specifications

Carrier GasHelium
Column Type30 m × 0.25 mm ID HP-5MS fused silica
Column Flow1.5 mL/min
Operating SystemMS Windows 95
ChemStation VersionProductivity ChemStation (revision A.03.00)

Frequently asked questions

What are the benefits of combining GC/MSD and GC/FID data?

The combined data provides identification with mass spectra, quantitation by the GC detector, faster method development, and maximum information from each analysis.

What is the recommended configuration for single column analysis?

The recommended configuration uses a single injection port and injector with the sample separated on a single column and split post-column to the two detectors.

How is the GC signal aligned with the MS signal?

A tool in the data analysis software allows the GC signal to be time aligned with the MS signal based on a two point linear fit.