Agilent HP-PLOT/Al2O3 Columns Application Note for C1-C6 Impurity Analysis

This article introduces the method of optimizing determination of C1-C6 impurities in propylene and ethylene using HP-PLOT/Al2O3 columns. Compared with other commercially available alumina PLOT columns, HP-PLOT columns have better selectivity, increased capacity and complete baseline separation. HP-PLOT column analyses were done using the 5890 Series II gas chromatograph with electronic pressure control and a flame ionization detector.


Agilent HP-PLOT/Al2O3 Columns Application Note for C1-C6 Impurity Analysis - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
147 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
Al2O3
MD5 checksum
1C01923D0A0E322726A17FFD9D2B37CB

About this manual

This document details the analysis of trace impurities in high-purity hydrocarbon gases, specifically propylene and ethylene. It compares the performance of Agilent HP-PLOT/Al2O3 columns against other commercially available alumina PLOT columns, such as Alumina/KCl and Brand X columns. The study highlights that HP-PLOT columns offer better selectivity, increased sample capacity, and complete baseline separation for critical impurities like cyclopropane, acetylene, and isobutane, which often suffer from masking or co-elution on other columns.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Results and Discussion
  • Analysis of Impurities in Propylene
  • Analysis of Impurities in Ethylene
  • Retention Index and Selectivity
  • Conclusion
  • References

Specifications

Gas Chromatograph Model5890 Series II with EPC

Frequently asked questions

How does the HP-PLOT column compare to Alumina/KCl columns?

HP-PLOT columns provide better selectivity and complete baseline separation, whereas Alumina/KCl columns often fail to separate critical impurities like cyclopropane from propylene.

What is the retention index requirement for acetylene quantitation?

For the quantitation of acetylene, the retention index (RI) should be greater than 400 to prevent coelution with isobutane.