Agilent 2100 Bioanalyzer Cell Fluorescence Assays Application Note

This Application Note describes the general use of cell fluorescence assays on the Agilent 2100 bioanalyzer. The cell assay extension together with the cell fluorescence LabChip® kit allows both the analysis of pre-stained cells and the analysis of cells by on-chip staining. Cell assays on the 2100 bioanalyzer are based on simple flow cytometric analysis. The technical principles, as well as the typical workflow in sample and chip preparation are described here. A typical dye optimization experiment was performed and the appropriate dye concentration for a specific cell line determined. The cell fluorescence software is easy-to-use and allows different forms of data evaluation and result presentation. Using an example, data evaluation is explained in detail. Specific advantages of cell fluorescence analysis with the Agilent 2100 bioanalyzer are low cell and reagent consumption, ease-of-use, automated data acquisition and evaluation of up to six samples.


Agilent 2100 Bioanalyzer Cell Fluorescence Assays Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
193 KB
Published
02 July, 2012
Updated
01 October, 2023
MD5 checksum
B619721680A7986517733DD13B1CF0B7

About this manual

The document details how to perform cell fluorescence analysis using the Agilent 2100 bioanalyzer with the cell assay extension and LabChip kit. It explains that the system uses simple flow cytometric analysis based on a two-color detection system (red laser at 635 nm and blue LED at 470 nm) to measure fluorescent labeled cells.

What's inside

  • Introduction
  • Prerequisites
  • Dye selection for cell staining
  • Cell lines
  • Instrument setup
  • Sample preparation
  • Data acquisition settings

Frequently asked questions

What is the recommended cell concentration for analysis?

A cell concentration of 2 million/ml is recommended to achieve a statistically relevant number of events.

How many samples can be analyzed per chip?

The kit allows for the analysis of six samples per chip.

What is the default sample run time?

The default sample run time is 4 minutes, which typically yields 500 to 1000 measured events.