Agilent Technologies Quality Control for the Agilent 2100 Bioanalyzer Protein 200 Plus LabChip Kits service manual

This technical note describes some of the typical quality control tests and specifications for Agilent 2100 Bioanalyzer Protein 200 Plus LabChip Kits.


Agilent Technologies Quality Control for the Agilent 2100 Bioanalyzer Protein 200 Plus LabChip Kits service manual - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
8
File format
PDF
File size
297 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
C2A95ECBAA8F1F041F5A99715C0BEC0A

About this manual

This document provides a technical overview of the Agilent Protein 200 Plus LabChip kits used for microfluidic protein analysis. It describes how the technology integrates sample handling, separation, staining, destaining, detection, and digital data processing to achieve high throughput and reproducibility.

The text details specific quality control tests performed on the kits, including sizing accuracy, resolution, detection limits, linear dynamic range, and area reproducibility. Results from various instruments and buffer conditions demonstrate that the kits meet factory specifications for protein analysis.

What's inside

  • Introduction
  • Protein 200 Plus Assay Specifications
  • Sizing Accuracy and Reproducibility
  • Sizing Resolution
  • Detection Limit
  • Linear Dynamic Range
  • Area Reproducibility

Specifications

Sizing range14-200 kDa
Typical sizing accuracy20% C.V.
Sizing resolution10% difference in MW
Minute detectable concentration (PBS)20 pg/mL BSA
Minute detectable concentration (0.5 M NaCl)40 pg/mL BSA
Samples per chip10 + ladder

Frequently asked questions

What is the typical sizing accuracy for the Protein 200 Plus assay?

The results are typically within 20% C.V. of the expected value.

What is the minimum detectable concentration for BSA in PBS?

The assay can detect 20 pg/mL of BSA in PBS with a good signal-to-noise ratio.

What is the specified linear dynamic range for the Protein 200 assay?

The linear dynamic range is specified for 0.02—2 mg/mL of protein.