Agilent Confirming the Presence of Drugs of Abuse in Samples Using an LC/Ion Trap Mass Spectrometer a Nearly 400 Drugs Compound Library Containing MS2 MS3 Spectra

In this work, a collision energy ramp is used in conjunction with an ion trap mass spectrometer to generate reproducible product ion spectra. For the purpose of identifying compounds in real samples, the generation of product ion spectra in those samples is also carried out using ramped collision energy. The process of confirmation includes the generating of compound spectra in the sample data, followed by NIST- based scoring of Fit, Reverse Fit, and Purity matching of library spectra. Searching of the MS spectra is not included because matrix interferences can result in poor scoring. Besides, the precursor ion spectra search is also not included because the search may result in overlapping with other compounds.


Agilent Confirming the Presence of Drugs of Abuse in Samples Using an LC/Ion Trap Mass Spectrometer a Nearly 400 Drugs Compound Library Containing MS2 MS3 Spectra - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
16
File format
PDF
File size
404 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
MS-3
MD5 checksum
FF177F59555BABB7C4B1CD599C1C123E

About this manual

The document details the use of an Agilent LC/MSD Trap XCT mass spectrometer to confirm the presence of drugs of abuse in samples. It utilizes a specific compound library containing MS2 and MS3 spectra for approximately 400 drugs and pharmaceuticals.

A key feature discussed is the SmartFrag technique, which involves ramping collision energy from 30% to 200% of a set value (1 V) over 40 milliseconds. This method generates reproducible product ion spectra by isolating precursor ions and fragmenting them via resonance excitation before smaller fragments can be further fragmented.

What's inside

  • Confirming the Presence of Drugs
  • LC/MSD Trap Mass Spectrometer
  • SmartFrag Technology
  • Ramped Collision Energy
  • MS2 and MS3 Spectra
  • Compound Library
  • NIST-based Scoring

Specifications

Collision energy range0.3 to 2.0 V (30% to 200%)
Set collision energy1 V
Ramp duration40 msec
Library sizenearly 400 drugs

Frequently asked questions

What technique is used to generate reproducible product ion spectra?

The SmartFrag technique, which involves ramping the fragmentation energy during the sequence.

What is the purpose of the library described in this document?

The library is used to confirm the presence of drugs of abuse and pharmaceutical compounds in samples.

How many drug compounds are included in the library?

The library contains almost 400 drugs of abuse and pharmaceutical compounds, with a specific list of 386 compounds provided.

What range of collision energy is used to generate the product ion spectra?

The analysis uses a ramped collision energy ranging from 30% to 200% of the set collision energy.

Download PDF (404 KB)