Agilent 1100 LC/MS Analysis of Simazine Thiobencarb Thiuram

This article describes a liquid chromatography/mass spectrometry method that has been successfully applied to the sensitive and simultaneous determination of the pesticides Simazine, Thiobencarb, and Thiuram. The method is selective and sensitive, making it suitable for the determination of these pesticides in drinking water.


Agilent 1100 LC/MS Analysis of Simazine Thiobencarb Thiuram - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
168 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
5985B7842D07ECE25E251B4467F2964A

About this manual

This document describes a liquid chromatography/mass spectrometry (LC/MS) method using electrospray ionization in positive ion mode for the simultaneous determination of the pesticides Simazine, Thiobencarb, and Thiuram. The method addresses the need for better selectivity and sensitivity compared to existing techniques, particularly for Thiuram.

What's inside

  • Background
  • Method
  • LC Conditions
  • Sample Analysis
  • Conclusion

Specifications

InstrumentAgilent 1100 Liquid Chromatograph/Mass Spectrometer
Ionization ModeElectrospray ionization positive ion mode
Drying Gas Flow8 L/min
Nebulizer Pressure40 psi
Fragmentor Voltage (Thiuram)40 V
Mass Range100-500 amu
Mobile Phase ACH3OH/30 mM CH3COONH4 (50/50)
Mobile Phase BCH3CH3
Gradient0 % to 100 % B in 20 min
Flow Rate0.2 mL/min
Oven Temperature40 °C
Injection Volume50 uL
ColumnInertsil ODS3, 3.1 mm id x 250 mm long x 5 um
Detection Limit (Simazine)0.02 ppb
Detection Limit (Thiuram)2.5 ppb
Detection Limit (Thiobencarb)1 ppb

Frequently asked questions

Which pesticides can be determined simultaneously?

The method is suitable for the simultaneous determination of Simazine, Thiuram, and Thiobencarb.

What are the detection limits for the target pesticides?

The detection limits are 0.02 ppb for Simazine, 2.5 ppb for Thiuram, and 1 ppb for Thiobencarb.

What instrument is used for this analysis?

The analysis uses an Agilent 1100 Liquid Chromatograph/Mass Spectrometer with electrospray ionization.

Download PDF (168 KB)