Agilent 6890/5973 GC/MSD BFB Tuning Application Note

This paper summarizes the standard method for tuning 4-bromofluorobenzene (BFB) and introduces three methods for tuning Agilent 6890/5973 GC/MSD systems.


Agilent 6890/5973 GC/MSD BFB Tuning Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
263 KB
Published
02 July, 2012
Updated
01 October, 2023
MD5 checksum
B296B9F8011161D3F83B4B3A282461A1

About this manual

This document outlines tuning criteria for USEPA methods 524.2, 8260B, and the Contract Laboratory Program Statement of Work (CLP-SOW) regarding 4-bromofluorobenzene (BFB) analysis. It details how the Agilent 6890/5973 gas chromatograph/mass selective detector system can be adjusted to meet these specific ion abundance requirements.

The text presents three distinct approaches to achieve a successful tune: using the standard automated procedure, modifying the autotune targets for improved sensitivity, and utilizing a specific 'Modified Autotune' procedure developed by the authors. It includes tables of criteria and screenshots of the EnviroQuant ChemStation software interface used for evaluation.

What's inside

  • Introduction
  • Tuning Criteria
  • Automated BFB Tuning
  • Target Tuning
  • Modified Autotune
  • Tune Evaluation

Frequently asked questions

What concentration of BFB standard was used for the manual injection shown in Figure 1?

The procedure involved a 1-uL injection of a methanol solution containing 50 ng/uL of BFB.

What is the maximum allowed relative abundance for m/z 174 under CLP-SOW requirements?

The CLP-SOW method allows the m/z 174 response to be up to 120% of the m/z 95 base peak.

Where is the BFB tune evaluation report archived by default?

The most recent report is saved in the datafile.d directory inside a file named tuneeval.txt.

Download PDF (263 KB)