Agilent 6890/5973 GC-MS System Application Note for EPA Method 524.2

This document describes a method for determining 61 volatile organic compounds (VOCs) using gas chromatography-mass spectrometry (GC-MS). The method uses a purge and trap (P&T) concentrator, a gas chromatograph (GC), and a mass spectrometer (MS), and does not require the use of cryofocusing. Tables 1 and 2 list the P&T and GC-MS conditions, which detail the instrument settings. The method development for analyzing 524.2 analytes was refined by members of the GC-MS Volatile Organics Analysis (VOA) group at Quanterra in Tampa, Florida.


Agilent 6890/5973 GC-MS System Application Note for EPA Method 524.2 - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
185 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
BFFB61316E0484439D0E68B08C59EDBE

About this manual

This document describes a system comprised of a purge and trap concentrator, an Agilent 6890 gas chromatograph, and an Agilent 5973 mass spectrometer used to determine 61 volatile organic compounds (VOCs). The method meets all U.S. EPA Method 524.2 criteria without using cryofocusing. The procedure involves purging compounds from water samples using helium, trapping them on a solid sorbent, and then desorbing them onto the head of a GC column for analysis by the mass spectrometer.

The document provides specific instrumentation and conditions for the purge and trap system (Tekmar LSC 3000) and the GC-MS system. It details the configuration of the Agilent 6890 GC, including the use of an EPC split/splitless inlet and a DB-624 column, as well as the operating parameters for the Agilent 5973 MS. The text also outlines the quality assurance procedures, calibration curves, and the typical instrument sequence for analysis.

What's inside

  • Introduction
  • Experimental
  • Purge and Trap Conditions
  • Gas Chromatograph and Mass Spectrometer Conditions
  • Instrument Sequence
  • Results

Specifications

Purge rate50 mL/min
Desorb temperature260 °C
Bake temperature270 °C
Inlet temperature200 °C
Split ratio35:1
Column length20m
Column diameter180 ym
Source temperature200 °C
Transfer line temperature250 °C

Frequently asked questions

How many volatile organic compounds can be analyzed using this method?

The system was used to determine 61 volatile organic compounds (VOCs).

Does this method require cryofocusing?

No, all U.S. EPA method 524.2 criteria were met without using cryofocusing.

What is the typical instrument sequence for analysis?

The sequence starts with a BFB tuning verification, followed by a CCV standard, LCS/LCSD, a method blank, and then field samples.