Agilent 6890/5973 Inert GC/MSD Semivolatiles Analysis Application Note

This document describes the application of the Agilent Technologies 6890/5973 inert gas chromatography/mass selective detector system in semivolatiles analysis. The system is designed for fast analysis while minimizing the impact of active compounds and maintaining linearity and sensitivity.


Agilent 6890/5973 Inert GC/MSD Semivolatiles Analysis Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
444 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
0B63E86BD318DBE283DA64349AFA1238

About this manual

This document describes the performance of the Agilent Technologies 6890/5973 inert gas chromatography/mass selective detector (GC/MSD) system for analyzing semivolatiles. The system is designed to meet the demand for lower minimum detection limits (MDLs) while maintaining linearity and sensitivity for active compounds.

The note details a specific method using a 100-um id column with a run time of 7.5 minutes and a calibration range of 5-200 ng. It compares the performance of the inert system against a standard Agilent 5973 system, highlighting improvements in response for active compounds and linearity across the calibration range.

What's inside

  • Abstract
  • Introduction
  • Instrument Operating Parameters
  • Results
  • Conclusions

Specifications

Total run time8.0 min
Column length12.5m
Column diameter100 um
Source temp230°C
Transfer line temp280°C

Frequently asked questions

What is the calibration range for the semivolatiles analysis?

The calibration range is 5-200 ng.

What is the total run time for the analysis?

The total run time is 8.0 minutes.

How many compounds are analyzed in the calibration standards?

The standards contain 77 compounds together with six internal standards.