Agilent Fast Analysis of Aromatic Solvent with 0.18mm I.D. GC column Application Note

A single, rapid GC method for aromatic solvent purity analysis is described. The application of a high-efficiency 0.18-mm-id GC column results in faster analyses compared to conventional 0.25-mm- or 0.32-mm-id columns without losing mea- surement performance.


Agilent Fast Analysis of Aromatic Solvent with 0.18mm I.D. GC column Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
16
File format
PDF
File size
1 MB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
529D5EB60303A82E26CCD051C04F2494

About this manual

This document details a method for the fast analysis of aromatic solvents using high-efficiency 0.18 mm internal diameter (ID) capillary columns. It addresses the need for increased sample throughput in QA/QC laboratories by reducing analysis time compared to conventional 0.25 mm or 0.32 mm ID columns without sacrificing measurement performance.

The note presents experimental results obtained using Agilent 6890N and 7890 gas chromatographs equipped with an Agilent 7683 Automatic Liquid Sampler. It compares three methods: a conventional method using a 60 m x 0.32 mm column, a method using a high-efficiency 20 m x 0.18 mm column, and an optimized fast analysis method. The document provides specific experimental conditions, including carrier gas pressure, inlet temperature, oven temperature programs, and detector settings, demonstrating that the optimized method achieves baseline separation for critical compound pairs like dodecane and o-xylene.

What's inside

  • Introduction
  • ASTM Methods for Aromatic Solvents
  • High-Efficiency Capillary GC Columns
  • Experimental Setup
  • Unified Aromatic Solvents Method (Method 1)
  • High-Efficiency Column Analysis (Method 2)
  • Fast Aromatic Solvents Analysis (Method 3)
  • Results and Discussion
  • Chromatograms

Specifications

Column TypeHP-INNOWax
Column Dimensions (Method 2)20 m x 0.18 mm x 0.18 um
Carrier GasHelium
Inlet Temperature250 °C
Detector TypeFID
Detector Temperature250 °C

Frequently asked questions

What are the advantages of using 0.18 mm ID columns?

They provide higher efficiency per unit of column length, allowing for shorter column lengths and faster analyses compared to conventional columns without losing measurement performance.

Which instruments were used for the aromatic solvent analysis?

The analysis was performed using one Agilent 6890N Series gas chromatograph and two Agilent 7890 gas chromatographs.

How does the optimized fast analysis method (Method 3) differ from the conventional method?

Method 3 uses a shorter 20 m x 0.18 mm column with a higher initial oven temperature (70 °C) and a faster heating rate (45 °C/min) compared to the conventional 60 m x 0.32 mm column.