Agilent Separation of Paraben Preservatives by Reversed-Phase HPLC Application Note

This document describes a rapid method for the analysis of paraben preservatives using reversed-phase HPLC.


Agilent Separation of Paraben Preservatives by Reversed-Phase HPLC Application Note - cover page
Brand
Agilent
Category
Other
Document type
Other
Language
English
Pages
4
File format
PDF
File size
61 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
D816B09A49DFA72EC826FC9BD6DD88A2

About this manual

This document details a method for the separation and analysis of paraben preservatives (methyl, ethyl, and propyl) and 2-phenoxyethanol (2PX) in foods, beverages, and cosmetics. The authors, Coral Barbas, Javier Rupérez, Andre Dams, and Ronald E. Majors, present a technique using reversed-phase HPLC to separate these compounds effectively.

The method utilizes a ZORBAX Eclipse XDB-C18 Rapid Resolution column with a mobile phase consisting of water and methanol. The analysis is performed at a flow rate of 0.8 mL/min and a temperature of 40°C, with detection at 254 nm. The document includes chemical characteristics, chromatographic conditions, and results demonstrating baseline separation within 16 minutes.

What's inside

  • Application
  • Introduction
  • Chemical Characteristics
  • Chromatographic Conditions
  • Results and Discussion
  • Conclusion
  • References

Specifications

Mobile phase AWater
Mobile phase BMethanol
Flow rate0.8 mL/min
Temperature40°C
DetectorUV 254 nm

Frequently asked questions

What column is used for separating paraben preservatives?

The method uses a ZORBAX Eclipse XDB-C18 Rapid Resolution column.

What is the analysis time for the paraben separation?

The analysis is completed in 16 minutes.

Download PDF (61 KB)

More Agilent Manuals