Agilent 1100 Series Normal Phase Analysis of Tocopherols in Margarine using HPLC Application Note

This document introduces a method for analyzing all eight naturally occurring tocopherols (T) and tocotrienols (T3) in margarine using normal-phase chromatography. This method uses fluorescence detection to avoid detecting peaks that overlap with coeluting peaks.


Agilent 1100 Series Normal Phase Analysis of Tocopherols in Margarine using HPLC Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
153 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
BE5FB97A578106EDD1BDB3ED7E2BE45B

About this manual

The document presents a high-performance liquid chromatography (HPLC) method for the analysis of tocopherols and tocotrienols in fats, oils, and foodstuffs. It highlights that normal-phase chromatography is required to separate all eight tocopherols completely, as reversed-phase chromatography cannot achieve full separation. The method utilizes a Hypersil SI 100 column with a mobile phase of hexane and 2% isopropanol.

What's inside

  • Abstract
  • Chromatographic conditions
  • Sample preparation
  • HPLC method performance
  • Equipment

Specifications

ColumnHypersil SI 100, 5 um
Mobile phasehexane + 2 % isopropanol
Stop time8 min
Flow rate0.3 ml/min
Column temperature25 °C
Injection volume0.5
Detector UV-DAD wavelength295/80 nm
Fluorescence emission wavelength330 nm
Sample preparation solventhexane
Repeatability (areas)<2%

Frequently asked questions

Can tocopherols be separated using reversed-phase chromatography?

No, tocopherols cannot be separated completely using reversed-phase chromatography.

What is the recommended sample preparation method?

Dissolve 20 g of sample in 15 ml of hexane.