Agilent 6890 Micro-ECD Analysis of Organochlorine Pesticides Application Note

A new electron capture detector (ECD) for the Agilent 6890 Series gas chromatograph (GC) was used to ana- lyze polychlorinated biphenyl congeners and organochlorine pesticides. The linearity of the 6890 Micro-ECD in the calibration range of 2 to 400 ppb was evaluated. The micro-ECD easily meets the linearity requirements of U.S. EPA contract laboratory programs for pesti- cides. Its limit of detection for these compounds goes down to less than 50 ppt. The micro-ECD also exhibits good reproducibility.


Agilent 6890 Micro-ECD Analysis of Organochlorine Pesticides Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
188 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
F88605B3C5BDB82189257927CF7DA5E4

About this manual

The document describes a new electron capture detector (ECD) developed for the Agilent 6890 Series gas chromatograph. This Micro-ECD features a smaller flow cell optimized for capillary columns, addressing limitations of traditional ECDs regarding linearity and sensitivity with small-bore columns.

What's inside

  • Introduction
  • Experimental Conditions
  • Linearity and Response Factors
  • Reproducibility
  • Detection Limit

Frequently asked questions

What is the linear calibration range for PCB congeners and organochlorine pesticides with the 6890 Micro-ECD?

The detector is linear for PCB congeners from 2 to 200 ppb and for organochlorine pesticides from 2 to 400 ppb.

What is the limit of detection for polychlorinated biphenyl congeners and organochlorine pesticides?

The limit of detection for these compounds goes down to less than 50 ppt.

What is the maximum allowable percent relative standard deviation for response factors in CLP methods?

The CLP method requires the percent RSD of the response factors to be less than 20 percent.