About this manual

This document details a quantitative proteomic study using the Agilent 6520 Accurate-Mass Q-TOF LC/MS System coupled with the 1200 Series HPLC-Chip/MS System. The study utilized 4-plex iTRAQ labeling to identify and quantify differentially expressed proteins in brain tissue samples from patients with tuberculous meningitis compared to normal controls.

The analysis involved sample preparation, fractionation by strong cation exchange chromatography, and LC/MS/MS analysis. The high resolution of the Agilent 6520 system allowed for the minimization of interferences in the iTRAQ reporter ion region (m/z 114-117), enabling reliable quantitation. Data analysis was performed using the Spectrum Mill MS Proteomics Workbench suite.

What's inside

  • Introduction
  • Sample preparation
  • iTRAQ labeling
  • Fractionation by strong cation exchange chromatography
  • LC/MS/MS analysis
  • HPLC-Chip conditions
  • Accurate-Mass Q-TOF LC/MS conditions
  • Data analysis
  • Results and Discussion

Specifications

Enrichment column160 nL
Analytical pump flow300 nL/min
Loading pump flow4 µL/min
MS mode acquisition rate1 spectrum/sec
MS/MS mode acquisition rate3 spectra/sec
Drying gas flow5 L/min
Capillary voltage1900 V
Collision energy slope3
Collision energy intercept2

Frequently asked questions

What software was used for protein identification and quantitation?

Spectrum Mill software was used for protein identification and quantitation of iTRAQ reporter ion intensities.

How many proteins were identified in the study?

Eight hundred proteins were identified from the LC/MS/MS analysis of the collected fractions.

What criteria were used to validate identified proteins?

Proteins were validated using a minimum peptide score of 6 and a protein score of 10 in the Spectrum Mill software.