Agilent 1290 Infinity LC Application Note for Aldehyde and Ketone Determination

This Application Note describes the development of a fast method for the determination of 13 aldehyde and ketone derivates with the Agilent 1290 Infinity LC system. The method, which used acetone as organic co-solvent separates the analytes within 3.5 minutes.


Agilent 1290 Infinity LC Application Note for Aldehyde and Ketone Determination - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
680 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
9B7A9D29575C0A5978811844AC386C0A

About this manual

This document details the development of a high-throughput method to separate 13 aldehyde and ketone derivatives within 3.5 minutes. The method utilizes the Agilent 1290 Infinity LC system, specifically leveraging its small dwell volume of 125 pL to enable fast separations with narrow bore columns. Acetone was used as the organic co-solvent in the mobile phase.

What's inside

  • Introduction
  • Experimental
  • LC system
  • Analyte mixture
  • Results and discussion
  • Conclusion

Specifications

Dwell volume125 pL
Maximum pressure1200 bar
Analysis time3.5 minutes
Flow rate1.2 mL/min
Detection wavelength360 nm
Column dimensions50 mm x 2.1 mm, 1.8 pm

Frequently asked questions

What is the analysis time for the described method?

The separation of 13 aldehyde and ketone derivatives is completed in around 3.5 minutes.

What organic co-solvent was used in the mobile phase?

Acetone was used as the organic co-solvent.

What is the critical resolution between peak pair 6 and 7?

The critical resolution is 1.6 between peak pair 6 and 7.