Agilent 1100 Series Sophisticated Peak-Based Fraction Collection Application Note

This Application Note describes a more sophisticated approach to peak-based fraction collection by using the parameters up and down slope, upper threshold and maximum peak duration.


Agilent 1100 Series Sophisticated Peak-Based Fraction Collection Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
380 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
BF685C2376962F9C94D48FEE68720C33

About this manual

This document explains how to perform sophisticated peak-based fraction collection using the Agilent 1100 Series purification system. It details the use of parameters such as up and down slope, upper threshold, and maximum peak duration to improve triggering decisions compared to standard threshold-only methods.

The text covers the theoretical basis of slope triggering, the function of the fraction preview tool in the Agilent ChemStation for optimizing settings, and provides practical applications for separating steep and shallow peaks. It also offers tips on setting reliable values for different chromatographic conditions.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Up and down slope
  • Fraction triggering decisions
  • Upper threshold
  • Maximum peak duration
  • Fraction preview tool
  • Separation of steep and shallow peaks
  • Applications

Frequently asked questions

What is the minimum up and down slope setting required for reliable fraction collection?

Reliable fraction collection cannot be expected with slope settings less than approximately 10 mAU/s.

What happens to the fraction collection trigger when the detector signal rises above the upper threshold?

Triggering decisions based on up and down slope are ignored until the signal falls back below the upper threshold.