Agilent Application Note: Separation of Sulfurated Amino Acids by HPLC with Pre-column Derivatization

This article introduces a method for separating 19 amino acids using an Agilent 1200 RRLC system, including two sulfurated amino acids, and transferring it successfully to an Agilent 1100 HPLC system. Good separation of this method was achieved by using Agilent ZORBAX Eclipse Plus C18 column on both two system. The results demonstrated that this analytical method was simple, reliable for analyzing these 19 amino acids, and the method is useful in comprehensively evaluating the quality of feed, food and pharmaceutical products which containing sulfurated amino acids.


Agilent Application Note: Separation of Sulfurated Amino Acids by HPLC with Pre-column Derivatization - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
168 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
2401DA0B2DC00A9546A564AD9114E609

About this manual

The document details a reliable analytical method developed to separate 19 amino acids, specifically addressing the challenge of analyzing sulfurated amino acids alongside standard ones. The method utilizes pre-column derivatization with OPA and FMOC reagents.

The procedure was initially developed using an Agilent 1200 SL RRLC system and successfully transferred to an Agilent 1100 HPLC quaternary gradient system. Both systems utilize ZORBAX Eclipse Plus C18 columns, demonstrating the method's versatility across different LC configurations.

What's inside

  • Abstract
  • Introduction
  • Experimental Equipment
  • Chemicals and materials
  • Chromatographic conditions
  • Pre-column derivatization Program
  • Results and discussion
  • Names and Order of Elution
  • Conclusions

Specifications

Column (1200 SL RRLC)46x50 mm, 1.8um Zorbax Eclipse Plus C18
Column (1100 HPLC)46x150 mm, 5 um Zorbax Eclipse Plus C18
Flow Rate (1200 SL RRLC)2 mL/min
Flow Rate (1100 HPLC)2 mL/min
Column Temp.40°C

Frequently asked questions

Which amino acids are analyzed in this method?

The method analyzes a total of 19 amino acids, which includes two sulfurated amino acids (Methionine sulfone and Cysteic Acid) along with seventeen others.

What reagents are used for pre-column derivatization?

The method uses OPA (Ortho-phthaldialdehyde) and FMOC (Fluorescein methyl ester carbamate) reagents.