Agilent LC/MS/MS quantitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note

This application note describes a sensitive, precise, and accurate quantitative liquid chromatography/tandem mass spectrometry (LC/MS/MS) method for the measurement of beta-estradiol acetate in human plasma. The sample preparation method employed solid phase extraction on Agilent SampliQ cartridges. The processed samples were chromatographed on an Agilent ZORBAX Eclipse Plus C18 analytical column and analyzed by triple-quadrupole tandem mass spectrometry in the multiple reaction monitoring (MRM) mode using negative electrospray ionization. The principal advantage of the LC/MS/MS method described in this application note is the simultaneous achievement of high absolute recovery (~ 80%), high sensi- tivity (LLOQ of 100 fg injected on-column or 28.6 pg/mL plasma), high precision (≤ 5.9% CV) and accuracy (81.5-118.1%), as well as excellent linearity over the concentration range 0.1 to 10000 pg/μL (R2 ≥ 0.998) with a short run time of only 1.2 min.


Agilent LC/MS/MS quantitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
177 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
E8EF004CD5D585D1742378F8257053C3

About this manual

This document details a sensitive and accurate LC/MS/MS method for measuring B-estradiol 17-acetate in human plasma. The method utilizes solid phase extraction with Agilent SampliQ cartridges and analysis via triple-quadrupole tandem mass spectrometry in negative electrospray ionization mode.

The study achieved high absolute recovery, sensitivity with a lower limit of quantification of 28.6 pg/mL, and excellent linearity over a concentration range of 0.1 to 10000 pg/pL with a short run time of 1.2 minutes.

What's inside

  • Abstract
  • Introduction
  • Chemicals
  • Preparation of standards
  • Human plasma extraction procedure
  • LC parameters
  • Triple-quadrupole parameters
  • Data acquisition and processing
  • Results and Discussion
  • Linearity tests

Specifications

Lower Limit of Quantification (LLOQ)100 fg injected on-column or 28.6 pg/mL plasma
Precision< 5.9% CV
Accuracy81.5-118.1%
Linearity Range0.1 to 10000 pg/pL
Run Time1.2 min
ColumnZORBAX Eclipse Plus C18 analytical column
Extraction CartridgeAgilent SampliQ C18 end capped (100 mg)
Flow Rate0.5 mL/min
Capillary Voltage-3500 V
Drying Gas Temperature300°C

Frequently asked questions

What is the lower limit of quantification for this method?

The method achieves a lower limit of quantification (LLOQ) of 100 fg injected on-column or 28.6 pg/mL plasma.

What is the run time for the analysis?

The analysis has a short run time of only 1.2 minutes.