Agilent 6460 Triple Quadrupole LC/MS Application Note for PFCA/PFSA Detection

This application note outlines a viable method for reliably detecting low-fem- togram levels of perfluorinated carboxylates (PFCA) and sulphonates (PFSA) without background component interference, with an inline contaminant trap LC modification. Good chromatographic resolution of all perfluorinated compounds was observed. A representative suite of PFCAs and PFSAs were analyzed herein and were all detected at on-column levels lower than 75 fg in drinking water matrices (S/N >3.) The most sensitive analytes PFHxS, PFDS and PFBS were detected at 2.6, 3.2 and 5 fg levels, respectively. Method detection limits for spiked pork liver matrix extract samples were below 600 fg on-column for the entire analyte suite. No detectable background conta- mination was observed in blank injections for any analyte in this study. Linearity for up to five orders of magnitude with R2 values above 0.996 for the entire suite were recorded.


Agilent 6460 Triple Quadrupole LC/MS Application Note for PFCA/PFSA Detection - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
465 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
FA5F7111F47388D0EF3EFDC87574D7EC

About this manual

This document outlines a viable method for reliably detecting low-femtogram levels of perfluorinated carboxylates (PFCA) and sulphonates (PFSA) using the Agilent 6460 Triple Quadrupole LC/MS system equipped with Agilent Jet Stream Technology. The method utilizes dynamic multiple reaction monitoring (dynamic MRM) and an inline contaminant trap configuration to achieve low limits of detection without background interference.

What's inside

  • Abstract
  • Introduction
  • Sample Handling
  • Instrumentation
  • Rapid Resolution HPLC Conditions
  • Mass Spectrometer Dynamic MRM Conditions
  • Ion Source Optimization
  • Results and Discussion
  • Inline Contaminant Trapping

Specifications

Column temperature55 °C
Autosampler temp4°C
Gradient flow rate0.5 mL/min
Total run time9.0 min

Frequently asked questions

What analytes were analyzed in this study?

The study analyzed a suite of perfluorinated carboxylates (PFCA) and sulphonates (PFSA), including PFBS, PFDA, PFDoA, PFDS, PFHpA, PFHxA, PFHxS, PFNA, PFOA, PFOS, PFTA, PFTriA, and PFUA.

What is the detection limit for PFHxS?

The most sensitive analyte PFHxS was detected at 2.6 fg levels.

What is the linearity range for the method?

Linearity was recorded for up to five orders of magnitude with R² values above 0.996.

What technology is used to increase sensitivity?

Agilent Jet Stream Technology was used, effectively increasing sensitivity to around 14x that of normal electrospray ionization (ESI) conditions.