Varian 210-MS Ion Trap GC/MS Application Note for EPA Method 525.2

This note provides a basic overview of 25 key regulated compounds rapidly set up on the 210-MS Ion Trap GC/MS system for the analysis of water samples by EPA Method 525.2.


Varian 210-MS Ion Trap GC/MS Application Note for EPA Method 525.2 - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
1.3 MB
Published
12 June, 2012
Updated
30 September, 2023
Model
210-MS
MD5 checksum
ED745C1767706BB252A6CA7E7E5AA45B

About this manual

This document provides an overview of analyzing 25 key regulated compounds for US EPA Method 525.2 using the Varian 210-MS Ion Trap Mass Spectrometer with a 431-GC Gas Chromatograph. It details the specific instrumentation, including the split/splitless injector and autosampler used in the setup.

The note outlines the initial calibration procedures, specifying concentration ranges from 0.05 to 10 ppm and acceptable %RSD limits. It presents detailed GC conditions (temperature programming) and MS parameters such as scan range, emission current, and ion trap temperature.

What's inside

  • Introduction
  • Instrumentation
  • Initial Calibration
  • GC Conditions
  • MS Conditions
  • Results and Discussion

Specifications

ColumnFactorFour VF-5ms 30 m x 0.25 mm x 0.25 pm
Injector Part NumberRT210462145
AutoSampler Model8400
Scan Range45-450 m/z
Emission Current15 pA
Ion Trap Temperature210 °C

Frequently asked questions

What is the acceptable %RSD for analytes in EPA Method 525.2?

The %RSD for each analyte should be less than 30% to use mean response factors or linear regression fitting.

Can quadratic calculations be used in EPA Method 525.2?

No, quadratic calculations may not be used according to the method requirements.

What is the concentration range for calibration solutions?

Calibration solutions of 0.05, 0.1, 0.2, 0.5, 1, 2, 5, and 10 ppm are used for all analytes.