This article describes the method for analyzing seven nitrosamines in the MS/MS configuration using the 220-MS Ion Trap GC/MS system and V:Results GC/MS software.
This document provides procedures for determining seven specific nitrosamine compounds (NDMA, NMEA, NDEA, NDPA, NDBA, NPYR, and NPIP) in finished drinking water using US EPA Method 521. The method utilizes a Varian 220-MS Ion Trap Mass Spectrometer coupled with a 450-GC Gas Chromatograph.
At the highest point, CAL standards must fall between 70-130% of their true value, and at the lowest point, they must fall between 50-150%.
This document provides procedures for determining nitrosamines in finished drinking water using the Varian 220-MS Ion Trap system.
The specified column dimensions are 60 m x 0.25 mm x 1.4um.