Agilent VF-17ms Column with EZ-Guard PAH Analysis Application Note

This document describes the GC analysis of PAHs using an Agilent J&W FactorFour VF-17ms column with EZ-Guard. The EZ-Guard is a protective device used to trap non-volatile components from the sample matrix to restore column performance. The document also demonstrates the selectivity of the VF-17ms column in analyzing these important pollutants.


Agilent VF-17ms Column with EZ-Guard PAH Analysis Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
136 KB
Published
12 June, 2012
Updated
28 September, 2023
MD5 checksum
412AD1CF01AF6CE3138D28EFEB09D3AE

About this manual

Polycyclic aromatic hydrocarbons (PAHs) are pollutants formed during incomplete combustion and pyrolysis. Legislation requires monitoring these substances in various matrices, which can introduce non-volatiles that diminish column performance.

The Agilent J&W FactorFour VF-17ms column is widely used for PAH analysis due to its high selectivity towards certain isomers. This document demonstrates the use of an EZ-Guard integrated with the column to trap non-volatile components, thereby extending column lifetime and maintaining resolution.

What's inside

  • Introduction
  • Peak Identification
  • Materials and Methods
  • Results and Discussion
  • Conclusion

Specifications

Column Dimensions30 m x 0.25 mm ID, df=0.25 pm
Guard Part NumberCP9025
Carrier GasHe
Injector Temperature275 °C
Detector TypeFID

Frequently asked questions

What is the purpose of using an EZ-Guard with the VF-17ms column?

The EZ-Guard traps non-volatile components from complex sample matrices to prevent them from adhering to the analytical column, thus restoring performance without losing resolution.

How does the EZ-Guard protect the analytical column from non-volatile components?

The EZ-Guard traps non-volatile components on its integrated guard, which prevents them from adhering to the analytical column and diminishing separation or detector response.

What is the procedure for restoring column performance when the EZ-Guard becomes saturated?

To restore performance, a meter of the EZ-Guard can be cut off, which allows the column to function again without losing resolution.

What are the dimensions and specifications of the FactorFour VF-17ms column used in this analysis?

The column used is 30 m long, has an inner diameter of 0.25 mm, and a film thickness of 0.25 µm.