Agilent GC Analysis of Base/Neutral Compounds using VF-17ms Column with EZ-Guard

This document describes the GC analysis of base/neutral compounds using an Agilent J&W FactorFour VF-17ms column with EZ-Guard. The EZ-Guard is an integrated guard column that optimizes system uptime and extends column lifetime. The document also provides specific experimental methods and results discussion.


Agilent GC Analysis of Base/Neutral Compounds using VF-17ms Column with EZ-Guard - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
118 KB
Published
12 June, 2012
Updated
28 September, 2023
MD5 checksum
5B2EC702466EFB73E5BC58A522063DAF

About this manual

This document details the use of the EZ-Guard integrated guard column to extend the lifetime of GC columns when analyzing complex matrix samples, such as those found in EPA 8270 methods. The EZ-Guard is designed to be leak-tight and inert, preventing coupling issues and active sites that can degrade peak shape.

The note provides specific experimental parameters for analyzing a mixture of base and neutral compounds, including column dimensions, carrier gas, temperature programming, and detector settings. Results demonstrate excellent peak shapes for various components, confirming the effectiveness of the EZ-Guard in maintaining column performance.

What's inside

  • Introduction
  • Materials and Methods
  • Results and Discussion
  • Conclusion

Specifications

ColumnFactorFour VF-17ms, 30 m x 0.25 mm, df=0.25 um
Guard Column10 m EZ-Guard (part number CP9025)
Sample Size1 uL
Split Ratio1:75
Sample Concentration20 yg/mL methylene chloride
Carrier GasH2
Pressure80 kPa
Oven Temperature Program50 °C to 300 °C (10 °C/min)
Injector Temperature275°C
Detector Temperature325°C
Detector TypeFID

Frequently asked questions

What is the benefit of using an EZ-Guard column?

The EZ-Guard is an integrated guard column that is leak-tight and has no active sites, which helps maintain peak shapes and high responses while extending column lifetime.

Which EPA method is mentioned for analyzing base and neutral components?

The document mentions EPA 8270 as a multicomponent environmental method for analyzing basic and neutral components.

What part number is assigned to the EZ-Guard used in this analysis?

The EZ-Guard used in this analysis has the part number CP9025.