Varian 210-MS GC/MS EPA Method 8260B Application Note

This document is an application note about the use of the Varian 210-MS ion trap mass spectrometer and V:Results™ GC/MS software for gas chromatography mass spectrometry analysis. The method is applicable for the determination of volatile organic compounds in various solid waste samples, including different water content samples such as air sampling trapping media, ground and surface water, aqueous sludges, caustic liquors, waste solvents, oily wastes, mousses, tars, fibrous wastes, polymeric emulsions, filter cakes, spent carbons, spent catalysts, soils, and sediments.


Varian 210-MS GC/MS EPA Method 8260B Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
605 KB
Published
12 June, 2012
Updated
28 September, 2023
Model
210-MS
MD5 checksum
3E6730050CAECA0993E59F5D65C4EFF6

About this manual

This document serves as an application note detailing the rapid setup of the Varian 210-MS Ion Trap Mass Spectrometer system for the analysis of volatile organic compounds (VOCs) according to EPA Method 8260B. It outlines the specific instrumentation configuration, including the Varian Archon Purge and Trap AutoSampler and V:Results GC/MS software.

The note provides detailed operating conditions for the purge-and-trap and gas chromatography components, as well as specific mass spectrometer parameters such as emission current and ion trap temperature. It presents calibration data and results for 60 volatile organic chemicals, demonstrating that the system meets or exceeds the performance criteria required by the method.

What's inside

  • Introduction
  • Instrumentation
  • Initial Calibration
  • Purge-and-Trap and GC Conditions
  • MS Conditions
  • Results and Discussion
  • Calibration Data
  • Conclusion

Specifications

ColumnFactorFour VF-624ms, 20 m x 0.15 mm x 0.84 um
Purge Volume5 mL at 40 mL/min
Split Ratio1:100
Target TIC12000 counts
Emission Current15-20 pA
Manifold Temp60 C
Transfer line Temp250 C

Frequently asked questions

What is the minimum correlation coefficient for calibration?

The minimum correlation coefficient required is 0.99.

What is the maximum allowed %RSD for calibration check compounds?

The %RSD of the calibration check compounds must be below 30%.