Agilent CP-Sil 88 for FAME GC Column Application Note

This document describes the separation of fatty acid methyl esters (FAME) on an Agilent J&W Select CP-Sil 88 column. The column has a high polarity liquid phase, which effectively differentiates between multiple FAME isomers.


Agilent CP-Sil 88 for FAME GC Column Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
133 KB
Published
12 June, 2012
Updated
28 September, 2023
MD5 checksum
2B0686C1B787F6D9DD88E5E68F67A988

About this manual

This document describes the routine separation of fatty acid methyl esters (FAME) using the Agilent J&W Select CP-Sil 88 column. The column utilizes a stabilized, highly substituted cyanopropyl siloxane phase to effectively separate positional FAME isomers based on small structural differences. The application demonstrates that all 37 components of a complex FAME mixture can be resolved in a single run using GC-FID.

What's inside

  • Introduction
  • Conditions
  • Peak Identification
  • Column Specifications
  • Oven Program
  • Injection Parameters
  • Detection Parameters

Specifications

Column Length100 m
Column Inner Diameter0.25 mm
Film Thickness0.2 um
Column Part NumberCP7489
Carrier GasHelium
Injection Temperature250 °C
Injection Volume0.5 yL
Detection Temperature270 °C

Frequently asked questions

How many FAME components were resolved in the test?

All 37 components were resolved in one run.

What type of detector was used for the separation?

A Flame Ionization Detector (FID) was used.

What is the part number for the CP-Sil 88 for FAME column?

The part number is CP7489.