Agilent J&W FactorFour VF-1701ms GC Analysis of Halogenated Pesticides Application Note

This application note demonstrates the performance of the VF-1701ms with EZ-Guard in the separation of pesticides.


Agilent J&W FactorFour VF-1701ms GC Analysis of Halogenated Pesticides Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
150 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
204717DF376DA2A3CA29ED94B12EABCD

About this manual

This document describes the use of the Agilent J&W FactorFour VF-1701ms gas chromatography column equipped with an EZ-Guard integrated guard column for analyzing halogenated pesticides. The EZ-Guard design eliminates leakage and adsorption risks associated with traditional guard columns, thereby improving column lifetime and sensitivity. The VF-1701ms is noted as a common conformal column for pesticide analysis, and this note highlights its effectiveness with the EZ-Guard system for complex sample matrices.

What's inside

  • Introduction
  • Materials and Methods
  • Results and Discussion
  • Peak Identification

Specifications

Column Dimensions30 m x 0.25 mm ID
Film Thicknessdf=0.25 um
Guard Column Length+ 10 m EZ-Guard
Guard Column Part NumberCP9177
Sample Size2 uL
Split Ratio1:50
Sample Concentration0.1% in hexane
Final Temperature220°C
Carrier GasH2
Carrier Gas Pressure90 kPa
Injector Temperature275 °C
Detector Temperature300 °C
Detector TypeECD

Frequently asked questions

What are the benefits of using EZ-Guard with the VF-1701ms column?

EZ-Guard eliminates the risk of leakage and adsorption, which decreases column lifetime and causes asymmetrical peaks, respectively.

What is the sample concentration used for the analysis?

The sample concentration was 0.1% in hexane.