Agilent 7890A Series GC System Application Note for VOC Analysis in Paints

This application note describes an analysis of volatile organic compounds (VOCs) in waterborne paints with an Agilent J&W DB-1701 GC column. The analysis was per- formed on an Agilent 7890 GC System and confirmed by Agilent GC/MS. The results indicate that the method is rapid, economic and reliable, and meets the requirements of VOC determination in waterborne interior architectural paint.


Agilent 7890A Series GC System Application Note for VOC Analysis in Paints - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
412 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
305D33A5FC409B2A8FBDD2CF24CB140E

About this manual

This document presents an analysis method for volatile organic compounds (VOCs) in waterborne interior architectural paints. The study utilizes an Agilent 7890 GC System equipped with an Agilent J&W DB-1701 column. The analysis is performed using both Gas Chromatography/Flame Ionization Detection (GC/FID) and Gas Chromatography/Mass Spectrometry (GC/MS) to ensure accurate identification and quantification of 24 specific VOC components.

What's inside

  • Introduction
  • Chemicals and standards
  • Sample preparation
  • Instrumentation
  • GC/FID Method
  • GC/MS Method
  • Results and Discussion

Specifications

ColumnAgilent J&W DB-1701, 30 m x 0.25 mm, 0.25 um
Carrier gasHelium, constant flow mode, 37 cm/s
GC Inlet temperature250 °C
Detector temperature (FID)260 °C
Internal standard2-butoxyethanol

Frequently asked questions

Which column is used for the VOC analysis?

The analysis uses an Agilent J&W DB-1701 column with dimensions of 30 m x 0.25 mm and 0.25 um film thickness.

What internal standard is used in the experiment?

2-butoxyethanol is used as the internal standard, while diethyl adipate serves as the internal boiling point marker.

Which systems are used for the analysis?

The experiments are performed using an Agilent 7890A Series GC System with an Agilent 7683 Injector and Sample Tray.